Proceedings on Automation in Medical Engineering
Vol. 1 No. 1 (2020): Proc AUTOMED
About monitoring wound size with time difference EIT
Main Article Content
Copyright (c) 2020 Proceedings on Automation in Medical Engineering
This work is licensed under a Creative Commons Attribution 4.0 International License.
Abstract
Electrical Impedance Tomography (EIT) is a non-invasive imaging technique, mainly applied to lung function monitoring. To determine its suitability to monitor healing dynamics of a wound, in a first step the time difference method (tdEIT) is applied to a cavity in a petri dish containing a thin layer of saline gelatin. The Position error (PE) and Area error (AE) for reconstructed tdEIT images were calculated for different frequencies and hyperparameter values. Evaluation of PE and AE shows that images reconstructed using hyperparameter range 0.005- 0.01 and higher frequencies produce more accurate results. Further studies are necessary with human 3D-tissue samples.